All tools / Calculators / Magnetic particle
Yoke lift test and ASME coil shot
How much a yoke has to lift under each document, and the ampere-turns and current ASME Section V asks for on a coil shot.
For reference only — verify against your procedure and the governing code.
Result
- Minimum lift
- 10 lb(4.5 kg)
- In newtons (converted from lb)
- 44.5N
At the maximum pole spacing you will use, with contact like the examination's.
How often: Before use each day the yoke is used, and after damage or repair (T-762(a)).
Source: ASME V T-762(b).
Head shot, internal conductor and prod current per inch are on the circular magnetization and prod current calculator.
Formula
- Lift: AC 10 lb (4.5 kg); DC 30 lb at 2–4 in. or 50 lb at 4–6 in. leg spacing (E1444, E709); DC or permanent magnet 40 lb (18 kg) (ASME V, ABS)
- ASME V coil, L/D ≥ 4: NI = 35,000 / (L/D + 2)
- ASME V coil, 2 ≤ L/D < 4: NI = 45,000 / (L/D); L/D < 2: no coil shot
- I = NI / N (within 10 %); L ≤ 18 in. (450 mm) per section; hollow: D(eff) = √(OD² − ID²)
Units. Lengths in mm or in. L/D is a ratio, so the coil formula is the same in both; only the 18 in. / 450 mm section length uses the figure ASME prints for each system. Leg spacings in mm are read against the 50/100/150 mm the documents print beside 2/4/6 in. Lift in lb and kg as printed, and newtons.
Assumptions and limits
- The lift figures are minimums at the leg spacing you will use, with contact like the examination's. ASME V checks before use each day; ABS at the start and end of each inspection day; E1444 every six months (Table 1); E709 suggests six months.
- E709 Table 3 lists the AC figure only under the 2–4 in. spacing, and E1444 and E709 give no figure for permanent magnets. Where a document gives no figure, this says so rather than borrowing another's.
- E709 prints 50 lb as 23.0 kg (225 N); E1444 prints it as 22.5 kg. Each is shown as its own document prints it; newtons elsewhere are converted from pounds (1 lbf = 4.448222 N).
- The coil formulas apply to direct or rectified current. Field adequacy is still shown with shims or a field indicator where ASME V asks for it (T-753.2(d), (e); T-764). With a Hall-effect probe on other techniques, ASME V T-764.2(c) looks for 30–60 G (2.4–4.8 kA/m) and E1444 §6.3.1.2 for at least 30 G.
Sources
- ASME Boiler and Pressure Vessel Code, Section V (2025), Article 7 — T-753.2(a)–(f) and its 10 in. × 2 in. example; T-753.3 and its 5-turn example; T-762 lifting power of yokes; T-764.2(c) tangential field.
- ASTM E1444/E1444M-26, Standard Practice for Magnetic Particle Testing — §7.4.4 dead weight check; Table 1 (every 6 months); §6.3.1.2 and A4.2 tangential field.
- ASTM E709-21, Standard Guide for Magnetic Particle Testing — §20.3.7 and Table 3 minimum yoke lifting force; Table 2 check interval.
- ABS Requirements for Nondestructive Testing (January 2025), Section 7 Magnetic Particle Testing — 5.7.2 magnetic field strength (test bar weights, daily checks).
- NIST Special Publication 811, Guide for the Use of the International System of Units, Appendix B conversion factors (lbf to N).
More magnetic particle calculators
For reference only — verify against your procedure and the governing code.
Standards and documents are cited by their publishers' names to identify them. ndt.support is not affiliated with, endorsed by or sponsored by ASTM International, ASME, ABS (the American Bureau of Shipping) or NIST. Formulas are paraphrased and cited by clause; get the documents themselves from their publishers.